The next step in characterization
The BELPORE mercury porosimeters precisely and quickly detect all ccessible pores
in the size range from 1 millimeter to 3.6 nanometers with tepwise equilibrium-controlled pressure increase
from a vacuum up to 414 MPa.
Accurate determination of the pore size distribution
The BELPORE mercury porosimeters from Microtrac MRB reliably and reproducibly masure pore diameters
from 1 millimeter to 3.6 nanometers at 414 mega-Pascals.
“Pascal” is not only a pressure unit, but also stands for the equilibrium-controlled and optimized handling
of the pressure build-up through the so-called “Pressurization by Automatic Speedup and Continuous Adjustment
Logic”, or P.A.S.C.A.L. for short.
This automatic control is regulated by the real pore system and permits shorter measuring times under guaranteed
equilibrium conditions as well as the detection of all pores within the specification – and this with up to 20,000
measuring points per analysis.
Since only three types of dilatometers are sufficient for all measurement tasks and neither gases nor liquid nitrogen
are required, running costs can be kept significantly low.
In addition, the BELPORE LP low-pressure porosimeter is easy to use and has an extended measuring range up to
pore sizes of 1 mm. The vertical degassing and filling with mercury on the BELPORE LP allows the degassing
pressure to be adjusted, making it possible to measure moist samples without changing the moisture content of
the material.
In this way, even moist concrete samples and solvent-containing porous green bodies can be measured unaltered.
The mercury porosimeters from Microtrac MRB are available for different pore ranges: *BELPORE LP (1,000 - 3.25 μm) *BELPORE MP (15 - 0.0065 μm) *BELPORE HP (15 - 0.0036 μm)
Model |
BELPORE LP |
Function |
Degassing, Hg filling, low-pressure porosimetry |
Pressure range |
Vacuum up to 450 KPa |
Resolution |
0.001 kPa in measuring range: vacuum to 0.1 kPA 0.01 kPa in measuring range 0.1 - 450 kPA |
Pressure detection accuracy |
Better than 0.1% F.S. |
Pore size diameter |
180 - 3.25 μm (UMP: 1,000 - 3.8 μm) |
Particle size diameter |
330 - 15 μm (UMP: 3,000 - 15 μm) |
Max. detectable volume |
0.5 cm³ - 2 cm³ |
Volume detection accuracy |
Better than 1% F.S. |
Max. number of measuring points |
10.000 intrusion 10.000 extrusion |
Weight & dimensions (W x D x H) |
55 kg (121 lbs) 40 x 67 x 80 cm |
Model |
BELPORE MP |
Function |
High-pressure porosimetry |
Pressure range |
0.1 - 228 MPa |
Resolution |
0.001 MPa in measuring range 0.1 - 100 MPa 0.01 MPa in measuring range 100 - 228 Mpa |
Pressure detection accuracy |
Better than 0.1% F.S. |
Pore size diameter |
15 - 0.0065 μm |
Particle size diameter |
40 - 0.015 μm |
Max. detectable volume |
0.5 cm³ - 2 cm³ |
Volume detection accuracy |
Better than 1% F.S. |
Max. number of measuring points |
10.000 intrusion 10.000 extrusion |
Weight & dimensions (W x D x H) |
68 kg (150 lbs) 40 x 67 x 80 cm |
Model |
BELPORE HP |
Function |
High-pressure porosimetry |
Pressure range |
0.1 - 414 Mpa |
Resolution |
0.001 MPa in measuring range 0.1 - 100 MPa 0.01 MPa in measuring range 100 - 414 Mpa |
Pressure detection accuracy |
Better than 0.1% F.S. |
Pore size diameter |
15 - 0.0036 μm |
Particle size diameter |
40 - 0.01 μm |
Max. detectable volume |
0.5 cm³ |
Volume detection accuracy |
Better than 1% F.S. |
Max. number of measuring points |
10.000 intrusion 10.000 extrusion |
Weight & dimensions (W x D x H) |
68 kg (150 lbs) 40 x 67 x 80 cm |
|