BELCAT II automated catalyst analyzer
측정항목
- Pulse chemisorption
- Temperature programmed desorption (TPD)
- Temperature programmed reduction (TPR)
- Temperature programmed oxidation (TPO)
- Temperature programmed reaction (TPReaction)
- BET single point measurement
특징
- Affordable price
- Small foot print
- Precise gas control by three high performance mass flow controllers
- Precise automatic gas injection system for calibration
- The pulse loop with temperature/pressure sensor enables the highly reproducible result
- A variety of options to fulfill the customers' demand
- With the standard gas mixing equipment, mixed gases (H2/Ar, O2/He, etc...) are not necessary
BELMASS II quadrupole mass spectrometer
The combination of the BELMASS II mass spectrometer with the BELCAT II makes it possible to evaluate the catalytic reaction (TPReaction)
in more detail. BELMASS ll can record the sample temperature and is suitable for thermal analysis.
Specifications
Measurement principle |
Dynamic flow method |
Detector |
Semi-diffusion type 4-element thermal conductivity detector (TCD) |
Measurement / pretreatment port |
1 |
Injection gas |
He, Ar, N2, O2, H2, CO, CO2, NH3, N2O, NO, etc. |
Dosing gas |
Carrier: 3 Pretreatment / pulse / MIX: 8 (corrosion-resistant x 2) |
Mass flow controller |
Carrier: F.S.100 sccm Pretreatment / pulse: F.S.100 sccm MIX: F.S.30 sccm |
Electric furnace |
Maximum temperature: 1,100°C Quick cooling: 30 minutes (400°C->50°) CATCryo II (option): -120°C supported |
Vapor injection (option) |
H2O, CH3OH, C2H5OH, toluene, benzene, etc. |
Dimensions (W x H x D) |
500 x 750 x 500 mm |
Weight |
80 kg |
Utility |
Gas: Measurement gas: 0.1MPa(G), valve driving gas: 0.45~0.55MPa(G) Joint: 1/8” Swagelok connection Power: AC 100-120 V / 200-240 V, 1,300 W, 50 / 60 Hz |
CE certified |
yes |
Recommended Monitor |
Full HD monitors |
|